molbloc-referenced gas flow, from sccm to thousands of slpm.

Our gas flow bench is built around the Fluke molbloc/molbox laminar flow measurement system, the same reference used by flow standards labs. Six molbloc elements cover 50 SCCM through 2,500 SLPM full scale, with uncertainty of ±0.25% of reading from 10–100% of each element’s range. Mass flow controllers are calibrated with live setpoint control and error mapping.
| Uncertainty (10–100% FS) | ±0.25% of reading |
|---|---|
| Below 10% FS | ±0.025% FS |
| Element coverage | 50 / 500 SCCM · 5 / 25 / 250 / 2,500 SLPM |
Reference standards in this lab
Fluke molbloc/molbox system
Laminar flow reference with six molbloc elements spanning 50 SCCM to 2,500 SLPM
Instruments we calibrate
- Mass flow controllers (MFCs)
- Mass flow meters
- Rotameters
- Flow meters
- Flow sensors
Representative list. The current A2LA scope PDF documents accredited ranges and CMC uncertainty at every point.