Pressure Calibration

Pressure up to 10,000 psi, at 0.01% of reading or better.

American Gage pressure calibration laboratory with rack-mounted pressure controllers, a hydraulic comparator bench, and nitrogen supply

The pressure laboratory covers pressure up to 10,000 psi, including full vacuum, with automated pressure controllers as references. Low-pressure work runs on a dedicated 0–30 inH₂O controller at ±0.009% of reading, pneumatic ranges to 2,500 psi at ±0.01% of reading, and hydraulic ranges to 10,000 psi.

Key specifications for Pressure Calibration
Range−14.7 to 10,000 psi
Pneumatic (to 2,500 psi)±0.01% of reading
Low pressure (0–30 inH₂O)±0.009% of reading
Hydraulic (to 10,000 psi)±0.01% FS per transducer range

Reference standards in this lab

  • Fluke 7250xi pressure controllers

    −14.7 to 2,500 psi pneumatic, ±0.01% of reading

  • Fluke 7615 hydraulic controller

    To 10,000 psi across 3,000 / 6,000 / 10,000 psi transducer ranges

  • Fluke (Ruska) 7250LP

    0–30 inH₂O low-pressure controller, ±0.009% of reading

Instruments we calibrate

  • Pressure gauges (analog & digital)
  • Pressure transducers & transmitters
  • Manometers
  • Digital pressure standards
  • Vacuum gauges
  • Barometers
  • Magnehelic & differential gauges

Representative list. The current A2LA scope PDF documents accredited ranges and CMC uncertainty at every point.